An AED that sits untouched still looks ready. The failure shows up at the worst moment. Here is what the data, a real court case and the law say happens when upkeep slips, and how to catch up fast.
The 30-Second Answer
Skipped maintenance leaves an AED with a dead battery, expired pads or an unnoticed fault. In a field test of 322 public AEDs, 21 percent failed at least one test phase. In an FDA report analysis, battery and power problems caused 23.2 percent of failures and pads and connectors caused 23.7 percent.
The legal cost follows the failure. Federal immunity for AED owners is conditioned on maintaining and testing the device. Start with the monthly AED check in five steps.
What This Article Covers
- What the failure data shows
- What actually fails in an unmaintained AED
- A real case: two AEDs that did not work
- Legal consequences of an unmaintained AED
- Who is responsible for AED maintenance
- Will your AED work with a low battery?
- What maintenance costs compared with skipping it
- The monthly AED check in five steps
- If you have skipped maintenance: a catch-up plan
- Frequently asked questions
1. What the Failure Data Shows
Two bodies of evidence answer the question. Neither gives a true failure rate, and both say so.
FDA failure reports, 1993 to 2008
A study in Annals of Emergency Medicine (2011) reviewed AED reports in the FDA's MAUDE database from 1993 through October 2008. Of 40,787 AED events, 1,150 were linked to fatalities.
- About 45 percent of failures happened while the device was charging or delivering a shock.
- Pads and connectors caused 23.7 percent of failures.
- Battery and power problems caused 23.2 percent.
- Only 4.7 percent of malfunctioning units showed a low-battery indication, and 9.6 percent shut off unexpectedly.
The caveats matter. The SCA Foundation notes that the data show association, not causation. Only 1.5 percent of reports included maintenance records, so it is unclear whether the failed devices were maintained. MAUDE is voluntary, so a failure rate cannot be calculated from it.
Separately, the FDA cited about 45,000 adverse event reports on AEDs between 2005 and 2012 when it moved to require premarket approval.
A field test of public AEDs
University of Louisville researchers tested 322 public AEDs at 190 sites in Seattle, Suffolk County (New York), Central Illinois and Louisville. They followed manufacturer guidelines for each unit.
- 21 percent failed at least one testing phase.
- 5 percent had expired batteries and would not power on.
- Registered AEDs passed more often. Seattle and Suffolk County, with registration above 80 percent, had 0 and 2 percent battery failures. Central Illinois, with registration under 25 percent, failed 38.2 percent of devices.
Sites volunteered, so the authors say the true risk is unknown. The finding that tracked, registered AEDs do better is the useful part. See AED registration requirements: state-by-state guide.
2. What Actually Fails in an Unmaintained AED
Most failures come from parts that wear out on a schedule. They are the parts a monthly check covers.
| Part | What goes wrong | The fix |
|---|---|---|
| Battery | It runs down or expires. An expired battery can leave the unit unable to power on. | Replace by the date. See the AED battery replacement schedule by brand. |
| Pads | The water-based gel dries, so expired pads stop working. | Replace before the printed date. See AED pad expiration: why pads expire and when to replace. |
| Cables and connectors | Cuts, loose connectors or a pad cable not seated. | Inspect the case and cables every month. |
| Software | Some recalls were fixed with a software upgrade, such as the 2009 ZOLL AED Plus action. | See AED software updates: when you need one and why. |
| Recalled units | A known defect with a manufacturer fix you never applied. | Check AED recall history and how to check your model. |
| Storage | Heat or cold pushes the AED outside its range. | Read AED heat and cold tolerance by model. |
| Never installed | The AED arrived without a battery and was never set up. | Open the box, install the battery, confirm the ready indicator. |
3. A Real Case: Two AEDs That Did Not Work
Diniro v. Aspen Athletic Club (2019 NY Slip Op 04858) shows how maintenance gaps turn into litigation. The facts below are allegations from the court record as reported, and the plaintiff must still prove them.
- The event. In May 2013 a member suffered cardiac arrest at an athletic club in DeWitt, New York. He died a month later.
- The older AED. It was inherited from the prior owners, had no low-battery warning, and its batteries were uncharged.
- The newer AED. It was delivered without batteries. The distributor had agreed to set it up, and discovery showed it never visited. The unit was still in its shipping box.
- The ruling. The court let the ordinary negligence claim against the club go forward, found no evidence for gross negligence, and dismissed both distributors because their actions did not cause the harm.
Two lessons follow. A device in the building is not the same as a working device, and the distributor who sold it did not carry the maintenance duty. The owner did.
The reported outcome is a procedural ruling, not a verdict. This is general information, not legal advice.
4. Legal Consequences of an Unmaintained AED
No federal law sets a maintenance schedule for AEDs. The exposure comes from the conditions attached to immunity and from ordinary negligence law.
| Source | What it says |
|---|---|
| Federal (42 U.S.C. 238q) | Acquirer immunity depends on notifying local emergency response, properly maintaining and testing the device, and training the user. It sets no inspection interval or record format. |
| California (HSC 1797.196) | Test at least biannually and after each use, inspect at least every 90 days, and keep records. See California AED laws and compliance. |
| Texas (HSC 779.003) | The owner or lessee must inspect the AED monthly. See Texas AED laws and compliance. |
| OSHA | OSHA's standards do not specifically address AEDs. Its guidance encourages them. |
Liability follows the cause of a failure. A manufacturing defect points toward the maker. Neglected maintenance, an expired battery or worn pads points toward the organization that owned the device.
That is why a device that fails from neglect can strip away the protection the owner was counting on. The Good Samaritan laws protect people who use an AED in good faith, which is a separate protection. Read Good Samaritan laws and AED use and liability and AED programs.
For the rules in your state, use AED laws and compliance by state. For the federal workplace picture, see OSHA AED requirements guide.
5. Who Is Responsible for AED Maintenance
The owner is. That is the person or organization that bought, leases or holds the AED.
- The owner or lessee. Texas puts the monthly inspection on the owner or lessee by statute.
- A named program coordinator. The American Heart Association AED guide assigns monthly checks to staff liaisons and records to the program.
- Not the seller. A distributor ships the device. Unless you buy a service plan, upkeep is yours.
Name one person and one backup. Put the task on a calendar. For multi-site help, see AED program management and ARCH AED medical direction and program management.
Who is liable if an AED fails during a rescue?
It depends on why it failed. A defect in the device points toward the manufacturer. A lapse in upkeep points toward the owner. A rescuer who followed the prompts in good faith is generally protected by Good Samaritan laws. A lawyer must apply this to your facts.
6. Will Your AED Work with a Low Battery?
Sometimes, and you should not count on it. A low-battery warning means the unit still has some power left, not that it is safe to ignore.
- ZOLL AED Plus. A red X appears at end of battery life, with about 9 shocks typically remaining.
- LIFEPAK CR2. The replacement indicator appears when at least 6 shocks and 30 minutes of operating time remain.
- Philips HeartStart FRx. A chirp with the Ready light off and the i-button flashing can mean a low battery, among other faults.
The warning is not guaranteed. In the FDA report analysis, only 4.7 percent of malfunctioning units showed a low-battery indication and 9.6 percent shut off unexpectedly.
Replace the battery when the indicator calls for it, or sooner, by the date. Read AED status indicators: decoding lights and beeps and AED self-test: what it checks and what failures mean.
If someone collapses, call 911 and use the nearest AED, even one that is overdue for maintenance. A shock from an imperfect AED is better than no shock. Follow the voice prompts and start CPR.
7. What Maintenance Costs Compared with Skipping It
Upkeep is small next to the device. Our complete AED maintenance guide puts ongoing upkeep (pads, battery, refreshers) at roughly $85 to $150 per year for a single unit, excluding the AED itself.
Over 10 years, a single-unit deployment runs approximately $4,000 to $5,500. Pad life runs 2 to 5 years by model and battery life 4 to 7 years. See AED total cost of ownership for four scenarios.
How much does it cost to replace expired pads?
It depends on the brand and model, so there is no single price. Browse the AED pads collection or the AED batteries collection, or call (855) 286-6384 to match the right pads to your model.
The cost of skipping is harder to price. It includes a failed rescue, the legal exposure described above, and the cost of a replacement unit if the AED is past saving. See When to Replace Your AED.
8. The Monthly AED Check in Five Steps
ZOLL, Defibtech and the American Heart Association all point to monthly checks. It takes about five minutes.
- Find it and read the indicator. The AED is in its place, visible and unlocked. The ready indicator is on, with no red light or chirp.
- Check the dates. Pads and battery are in date.
- Inspect the case and cables. Look for cracks, cuts and loose parts.
- Check the cabinet and signs. The door latches, the alarm works and signs are visible. See when to replace AED cabinets: wear, ADA updates and aesthetics.
- Sign and log it. Date the Avid Safety CPR AED inspection tag and record the result in your AED maintenance logs: what to document and why.
Do not switch the AED on to test it. The indicator already reports the last self-test. For the full printable list, use the AED inspection checklist: monthly, quarterly and annual.
Can I automate my AED maintenance tracking?
Yes, in part. The AED runs its own self-tests, but they do not replace your monthly check. Automate the reminders and the records.
- One AED: a recurring calendar reminder tied to the printed dates is enough.
- Several AEDs: keep one register with each unit's model, serial number, battery and pad dates and last inspection.
- Many sites: tracking software runs roughly $50 to $200 per AED per year, and a service program can take over checks. See AED program management.
9. If You Have Skipped Maintenance: A Catch-Up Plan
Do not wait for the next scheduled check. Do this today.
- Read the indicator. If it shows a fault, note the pattern and call the manufacturer.
- Read the dates on the pads and battery. Replace anything expired. Order from the AED pads collection and the AED batteries collection.
- Check for a recall. Use the serial number on the back of the unit. See five steps to check an AED recall.
- Check where it lives. Move it out of heat, cold or direct sun.
- Start the log. Record today's date and what you found. Do not backdate old checks.
- Confirm your registrations. Make sure EMS or your state registry knows where the AED is.
- Decide on replacement. If the unit is past its service life, plan a new one. Compare options in AED value packages.
If you are retiring a unit, follow how to dispose of an old defibrillator.
10. Frequently Asked Questions
What happens if an AED is not maintained?
Batteries run down, pads dry out and faults go unnoticed. In a four-region field test, 21 percent of public AEDs failed at least one test phase. The owner also risks losing the legal protection that depends on maintenance. See what the failure data shows.
How often should an AED be checked?
Monthly is the common standard, set by ZOLL, Defibtech and the American Heart Association guide. Texas requires a monthly inspection, and California requires inspection at least every 90 days.
Why is it important to check AED pads and batteries?
Together they caused about 47 percent of the failures in the FDA report analysis: 23.7 percent for pads and connectors and 23.2 percent for battery and power. Both are on a printed expiration clock.
Why do AED pads have an expiration date?
The conductive gel is water-based and dries out over time, and dried pads will not work. Sealed packaging does not stop the clock. See AED pad expiration: why pads expire and when to replace.
Who is responsible for AED maintenance?
The owner or lessee. Texas states this by law, and the American Heart Association guide assigns monthly checks to a staff liaison. The seller is not responsible unless you buy a service plan. See who is responsible for AED maintenance.
What are the legal requirements for AED maintenance?
Federal law sets no schedule, but 42 U.S.C. 238q conditions owner immunity on maintaining and testing the device. States add specifics, such as California's 90-day inspection and Texas's monthly inspection. See AED laws and compliance by state.
What are the legal consequences of an unmaintained AED?
The owner can lose the immunity that depends on maintenance and face a negligence claim if the AED fails because of neglect. Penalties vary by state. This is general information, not legal advice.
Who is liable if an AED fails during a rescue?
It depends on the cause. A manufacturing defect points to the manufacturer, and neglected upkeep points to the owner. A good-faith rescuer who followed the prompts is generally protected. See liability and AED programs.
Will my AED work if the battery is low?
Often for a limited number of shocks, but not reliably. A ZOLL AED Plus typically has about 9 shocks left when its red X appears, and a LIFEPAK CR2 warns at 6 shocks or 30 minutes remaining. In the FDA analysis, few failed units showed a low-battery warning at all.
How much does it cost to replace expired pads?
It varies by brand and model. Ongoing upkeep averages roughly $85 to $150 per year for one unit. See the AED pads collection or call (855) 286-6384.
How do I perform a monthly AED check?
Read the indicator, check the pad and battery dates, inspect the case and cables, check the cabinet and signs, then sign and log it. Do not power the AED on to test it. Follow the monthly AED check in five steps.
Can I automate my AED maintenance tracking?
Yes. Use a calendar reminder for one AED, a register for several, or tracking software or a service program for many sites. The AED's own self-tests do not replace your monthly check.
Does the AED's self-test replace monthly checks?
No. The self-test checks the battery, pads and circuits and reports through the indicator, but only a person sees a cracked case, a blocked cabinet, an expired pad date or a dead indicator. See AED self-test: what it checks and what failures mean.