AED Self-Test: What It Really Tests and What It Misses

Every modern AED tests itself automatically, and a green light tells you the last test passed. Here is what each brand's self-test checks, how often it runs, and the gaps that still need a person.

Reading time: 7 minutes Last updated: October 2026 Reviewed by: Kyle Nowack, BLS Instructor Trainer

The 30-Second Answer

An AED self-test automatically checks the battery, the pads and the internal circuits on a fixed schedule: daily on Philips, LIFEPAK CR2, Cardiac Science and Defibtech, every 7 days by default on the ZOLL AED 3, and weekly on HeartSine. It reports the result on a status indicator. It does not check for physical damage, spare pads, location, cabinet condition, or dates that some models do not read, so a monthly inspection is still required. A failed test means stop and follow your manufacturer's steps in Section 4.

A passing self-test proves only what the device can measure about itself. This article sets out what each major brand tests, based on manufacturer documents, and what the AED inspection checklist: what to check monthly covers instead.

1. What Each AED Self-Test Checks, by Brand

Manufacturers publish different levels of detail, so the table shows what each one states. A blank means the document does not say, not that the AED skips the check.

AED How often What the manufacturer says it checks Passed result
Philips HeartStart FRx and OnSite Daily, weekly and monthly Internal circuitry, waveform delivery system, pads (including gel moisture) and battery capacity Blinking green Ready light
ZOLL AED 3 Every 7 days by default, or daily; monthly 200 J test Battery capacity, status and expiration; electrode connection and expiration; ECG and charge circuits; processor hardware and software; CPR circuitry; audio Green check
ZOLL AED Plus Ongoing internal self-checks Not stated in the documents we reviewed Green check; a red X signals a problem
Stryker LIFEPAK CR2 Daily, weekly, monthly and at power-on Electrodes (connected, used, expired); alerts on any readiness issue Green light flashes every 6 seconds
HeartSine samaritan PAD Weekly, Sunday midnight GMT Components not listed in the manual Green light flashes every 5 to 10 seconds
Defibtech Lifeline Daily, weekly, monthly and quarterly circuitry tests Circuitry, pre-connected pads (daily), battery system test on insertion Flashing green when ready
Cardiac Science Powerheart G5 Daily, weekly and monthly Battery, pads, electronics and buttons daily; adds CPR accelerometer and high-voltage circuits weekly Green Rescue Ready indicator

If a self-test points to a battery or pad problem, the fix is a like-for-like part: the Philips M5070A AED replacement battery (4-year), the ZOLL AED 3 non-rechargeable lithium battery, or the Physio-Control LIFEPAK CR2 4-year lithium battery. Pads come from the AED pads collection.

2. Daily, Weekly and Monthly Tests

The tests escalate. On the Cardiac Science G5, the daily test checks the battery, pads, electronics and buttons, the weekly test adds the CPR accelerometer and high-voltage circuits, and the monthly test adds full-energy high-voltage tests. Defibtech adds a quarterly circuitry test, the ZOLL AED 3 runs a monthly 200 joule test, and HeartSine runs one weekly test.

3. What the Self-Test Misses

The self-test reports on the device. It cannot see its surroundings or the items kept beside it. These gaps are why the AHA's AED guide assigns a person to check every AED monthly.

  • Physical damage. HeartSine's manual calls for a monthly look for damage in addition to the weekly indicator check.
  • Dates the AED does not read. HeartSine tells owners to read the Pad-Pak date monthly, and Philips' manual describes a gel moisture test but no date check. Our guide to AED pad expiration: why it matters and when to replace shows which models flag expired pads.
  • Spare pads and the rescue kit. The AHA recommends two unexpired adult pad sets with each AED, plus pediatric pads where children are present.
  • Location, access and signs. The AHA says not to lock an AED away and to inspect signs at least yearly.
  • Results that never reach you. A LIFEPAK CR2 that is not connected to Wi-Fi still self-tests but sends no email.
  • Recalls and notices. Check the FDA's medical device recall database for your model. See AED recall history and how to check your model.

4. If the Self-Test Fails or the AED Beeps

A failed self-test announces itself by light, sound or both. The response is brand specific, and our guide to how to read AED status indicators: lights and beeps covers each pattern in full.

  • Philips FRx. A missing Ready light with single chirps and a flashing blue i-button points to a self-test error, pads, a low battery or a left-in Infant/Child Key. Triple chirps mean call Philips. Our Philips FRx SMART Pads II defibrillation electrode pads are the replacement when pads are the cause.
  • ZOLL AED 3. A blank status window means the last self-test failed. Confirm the pad cable, replace the pads, hold On/Off for 5 seconds to run a self-test, and if it still fails, remove the AED from service and call ZOLL Technical Service.
  • HeartSine. For flashing red, continuous beeping or no light, check the Pad-Pak date first. If it is in date, press On/Off and listen for "Call for medical assistance," then call HeartSine at 800 442 1142. A replacement is the HeartSine samaritan Pad-Pak electrodes and battery cartridge.
  • LIFEPAK CR2. The green flashing light stops and an alert sounds, and a Wi-Fi unit emails its administrator. Stryker says to follow up as soon as possible.
  • Defibtech Lifeline. Flashing or solid red, no flash, or a dark indicator means service. Turn the AED on and listen to the prompts, and run a manual self-test if it was dropped.
In an Emergency

If someone has collapsed and the only AED is beeping, call 911, start CPR, and try the AED. If it powers on and gives voice prompts, follow them. If it stays dead, keep doing compressions until help arrives. Afterward, take the AED out of service until the fault is fixed.

5. Temperature and Self-Test Failures

Heat and cold can fail a self-test. HeartSine's AED gives three rapid beeps after the weekly test when the ambient temperature is outside 0 to 50 degrees Celsius (32 to 122 degrees Fahrenheit), and the fix is to return it to range. In 2025 the FDA recorded a Class 2 recall of 73 ZOLL Powerheart G5 units (Z-1320-2025) that failed self-tests after extreme temperature or humidity. Keep an AED between 0 and 50 degrees Celsius, outdoors only in a temperature-controlled AED wall cabinet, and see AED heat and cold tolerance by brand.

6. Alerts and Records

A Wi-Fi LIFEPAK CR2 emails designated recipients when a self-test finds an issue and updates its status in LIFELINKcentral, and fleets can track every unit with AED program management software. Stryker advises against opening a CR2 lid after the first inspection except in an emergency, because opening the lid reduces battery power. Keep a dated record of each monthly check, because federal law (42 U.S.C. 238q) withholds acquirer immunity for failure to properly maintain and test an AED. Our guide to AED maintenance logs and what to document shows what to record.

7. Frequently Asked Questions

How do I know if my AED is working?

Read the status indicator, which shows the result of the last self-test. Then do the monthly inspection for what the self-test cannot see: damage, pad and battery dates, spare pads and location. See Section 3.

Why is my AED beeping?

A beep or chirp signals a fault: a self-test error, low battery, pad problem, or temperature outside range. The pattern differs by brand, so use Section 4 and your manual.

How often do AEDs perform self-tests?

Daily on Philips, LIFEPAK CR2, Cardiac Science and Defibtech, every 7 days by default on the ZOLL AED 3, and weekly on HeartSine. Most also run longer weekly and monthly tests.

What happens if an AED self-test fails?

The indicator changes and the AED beeps, and the AED is not rescue-ready. Follow your manufacturer's steps, replace an expired or depleted part, and call the manufacturer if the fault stays.

What does the status indicator on an AED mean?

It shows the last self-test result: a green check on ZOLL, a flashing green light on Philips, LIFEPAK CR2, HeartSine and Defibtech. Red, dark or blank means a problem. See our guide to AED status indicators: lights and beeps.

If the AED is beeping, can I still use it in an emergency?

Try it. Call 911 and start CPR, and if the AED powers on and prompts, follow it. If it does not respond, continue CPR. Take it out of service afterward.

Does the self-test drain the battery significantly over time?

No manufacturer publishes the share of battery the tests use, and published standby lives apply to installed AEDs running their normal tests. What cuts life is extra use, such as opening the lid, so leave it closed. See the AED battery replacement schedule by brand.

How do I document a passed self-test for legal liability protection?

Record the date, the indicator you saw and your initials at each monthly check, and keep the log with the AED. Federal law withholds acquirer immunity for failure to maintain and test, and state laws differ, so confirm with counsel. See Section 6.

What does the AED check during a daily self-test, and how do daily, weekly and monthly tests differ?

Philips checks circuitry, waveform delivery, pads and battery capacity daily. Each longer test repeats the shorter one and adds more, such as full-energy high-voltage tests monthly on the G5. See Section 1 and Section 2.

Is there a way to get remote notifications if a self-test fails?

Yes. A Wi-Fi LIFEPAK CR2 emails designated recipients, and program management software tracks fleets. A non-connected AED sends nothing, so check its indicator.

What should I do if the status indicator is blank?

Treat it as a failure. On the ZOLL AED 3 a blank window means the last self-test failed, and on a Philips FRx no light and no chirp means the battery is missing or depleted. Check the battery, then call the manufacturer.

Can extreme temperatures cause a self-test to fail?

Yes. HeartSine gives three rapid beeps outside 0 to 50 degrees Celsius, and a 2025 ZOLL Powerheart G5 recall covered self-test failures after extreme temperature or humidity. See Section 5.

See our full shipping policy and return policy before ordering, or call (855) 286-6384 to match batteries and pads to your exact AED model.

Kyle Nowack, BLS Instructor Trainer and AED Program Specialist at Total AEDs
Written and Reviewed By

Kyle Nowack

BLS Instructor Trainer, AED Program Specialist

Kyle Nowack is a certified American Heart Association BLS Instructor Trainer and AED program specialist at Total AEDs. His work focuses on helping workplaces, schools, and public-access programs build defensible AED deployments, from initial device selection through compliance documentation and responder training. Kyle has trained hundreds of workplace and school responders on CPR and AED use across the U.S.

Add Comment

?
TOP
0 Items
Logo