What an AED program manager needs to keep every defibrillator rescue-ready: what to check and when, how pad and battery life works by brand, how to track recalls, and when to retire a unit.
The 30-Second Answer
An AED stays rescue-ready through four routines: a monthly readiness inspection, pad and battery replacement by the printed expiration date, tracking of software updates and recalls, and written records of all three. Pads last 2 to 5 years and batteries 4 to 7 years depending on the model, and used pads are replaced immediately after every rescue.
Some states write intervals into law. California, for example, requires every AED to be inspected at least every 90 days and tested after each use. A monthly routine meets that floor.
An AED can sit untouched for years and still be expected to work on the first try. The FDA notes that survival probability drops 7 to 10 percent for every minute a victim stays in a life-threatening arrhythmia, so a dead battery or dried-out pads cost the one thing a rescue does not have: time. The failure is rarely the device. It is usually a routine nobody owned, as the real-world consequences of skipped AED maintenance show. Each section below summarizes what matters and links to a deeper article.
What This Guide Covers
- The 4 maintenance categories every AED needs
- Daily and weekly visual checks
- Monthly inspection checklist
- Pad lifecycle and replacement
- Battery lifecycle and replacement by brand
- Software updates and manufacturer notices
- Documentation and program records
- When an AED reaches end of life
- Recall history and how to check
- Frequently asked questions
Figures come from manufacturer data sheets (Stryker, ZOLL, Defibtech, Philips), the American Heart Association AED implementation guide, FDA recall records, and the cited statutes. Where retailers disagree with a manufacturer, the manufacturer's number is used.
1. The 4 Maintenance Categories Every AED Needs
Every AED program, from one unit in a break room to a fleet across campuses, runs on the same four routines.
Readiness inspections
Visual checks of the status indicator, case, cables, location, and rescue kit. An owner task that takes minutes.
Consumables
Pads and battery replaced by their printed dates and after use. The most common cause of AED non-readiness.
Software and recalls
Firmware updates and manufacturer or FDA notices, which reach you only if the AED is registered.
Records
Inspection logs, expiration dates, serial numbers, EMS registration, and post-use reports.
Every current AED runs automatic self-tests. The LIFEPAK CR2 tests daily, the HeartSine samaritan PAD 350P and 360P test weekly, and the ZOLL AED 3 tests every 7 days by default (configurable to daily) with a monthly full-energy test at 200 joules. A self-test checks the electronics, battery, and pads, but not a locked cabinet, a missing rescue kit, or a moved AED. See AED self-test: what it tests and what it misses.
2. Daily and Weekly Visual Checks
The documented US baseline is a monthly inspection, but the cheapest safeguard is a habit: anyone walking past glances at the status indicator. The American Heart Association's guide tells employees to confirm the indicator shows ready and to report damaged or missing units, a blinking red light, or a chirping sound.
- Daily glance (anyone). Status indicator shows ready, no chirping, cabinet closed and in place. Ten seconds, no tools.
- Weekly walk-by (assigned person). The British Heart Foundation recommends weekly physical checks. Public-access, outdoor, and high-traffic locations gain the most, because a failure there sits unnoticed longest.
Do not switch the AED on to test it. The status indicator already reports the most recent self-test, and powering up only drains the battery. If the indicator shows a fault, report it the same day and call the manufacturer. how to read AED status indicators: lights and beeps covers each brand's signals, and when to replace a worn AED cabinet covers the housing.
3. Monthly Inspection Checklist
ZOLL, Defibtech, and the American Heart Association AED implementation guide all set monthly as the inspection interval. Use the manufacturer's checklist if one exists; this list covers the items they share.
- Location and access. The AED is in its designated spot, visible, unobstructed, and not locked away. The AHA warns that locking an AED in an office or cabinet can delay use.
- Status indicator. Ready indicator on, no red light, no chirping.
- Case, cables, and connectors. Clean and dry, no cracks or loose parts, cables free of cuts and exposed wire.
- Pads. Present, sealed, and in date. The AHA recommends two sets of unexpired adult pads in every storage box, plus pediatric pads where children are present.
- Battery. Within its expiration or install-by date. If your AED uses disposable batteries, keep a fresh backup on hand.
- Rescue kit. Pocket mask with one-way valve and scissors present.
- Environment. Stored within the manufacturer's range, 32 to 122 degrees Fahrenheit (0 to 50 degrees Celsius) for most current models.
- Signage and sign-off. Signs visible and lit (the AHA says to inspect signs at least once a year). Sign and date the inspection tag and send a copy to the program coordinator.
Total AEDs publishes the full printable version in AED inspection checklist: what to check monthly, and the AED heat and cold tolerance guide covers sites with extreme climates.
4. Pad Lifecycle and Replacement
Pads are single-use and replaced immediately after any use on a patient. Unused pads expire too: the conductive gel is water-based and dries out over time, and dried pads will not work. Sealed packaging does not pause the clock, and the AHA notes that extreme temperatures degrade pads faster. Keep the package sealed until an emergency. On pre-connected designs such as the Philips FRx, the pads stay plugged into the AED with the foil intact.
Shelf life is set by the model, so read the printed date rather than assuming a cycle:
- 2 years. Philips HeartStart pads, Defibtech Lifeline adult AED pads, and Cardiac Science Powerheart G5 pads.
- 4 years. LIFEPAK CR2 QUIK-STEP electrodes, and the HeartSine Pad-Pak, which pairs pads and battery under one expiration date.
- 5 years. ZOLL AED Plus CPR-D-padz and ZOLL AED 3 CPR Uni-padz.
Pads are also model-specific. Philips OnSite and Home pads do not fit the FRx, which uses its own Philips FRx SMART Pads II defibrillation electrode pads, and Cardiac Science G3 and G5 pads are not interchangeable. AED pad expiration: why it matters and when to replace explains the gel science and replacement timing.
5. Battery Lifecycle and Replacement by Brand
A battery rating is a standby life that starts at installation and assumes the AED's routine self-tests. Batteries also carry a shelf life before installation. The Philips M5070A, for example, is rated for 4 years of standby once installed and 5 years on the shelf, with an install-by date printed on the label. The LIFEPAK CR2 battery delivers its 4-year standby life when installed within a year of manufacture.
| AED | Battery standby life | Replacement |
|---|---|---|
| ZOLL AED Plus | 5 years | ZOLL AED Plus replacement lithium batteries (10-pack) |
| ZOLL AED 3 | 5 years once installed | ZOLL AED 3 non-rechargeable lithium battery |
| Stryker LIFEPAK CR2 | 4 years, if installed within a year of manufacture | Physio-Control LIFEPAK CR2 4-year lithium battery |
| Philips HeartStart OnSite, Home, FRx | 4 years installed, 5 years on the shelf | Philips M5070A AED replacement battery (4-year) |
| HeartSine samaritan PAD 350P and 360P | Pad-Pak: battery and pads in one cartridge, 4 years from manufacture | HeartSine samaritan Pad-Pak electrodes and battery cartridge |
| Defibtech Lifeline and Lifeline VIEW | Lifeline: 5-year or 7-year pack. Lifeline VIEW: 4-year pack | Call (855) 286-6384 |
| Cardiac Science Powerheart G5 | 4-year operational guarantee | Cardiac Science Powerheart G5 replacement battery |
Replace the battery after a rescue if the manual calls for it, since some batteries must be changed after any use and others handle multiple rescues. The AED battery replacement schedule by brand carries the complete chart with capacities and part numbers. Every replacement is in the AED batteries collection and the AED pads collection.
6. Software Updates and Manufacturer Notices
Newer connected AEDs update themselves. Stryker lists automatic daily self-tests, software updates, and status notifications for the LIFEPAK CR2. Older platforms need a manual update or a service visit, so ask the manufacturer whether your model still receives software support. When an AED needs a software update explains how to tell.
Updates and recall notices only reach you if the manufacturer knows where the AED is. The AHA guide says to register each AED with its manufacturer. Manufacturer platforms then do the tracking: ZOLL PlusTrac monitors consumable expiration, training dates, and inspections, and alerts customers to device notices and recalls. Defibtech offers DefibtechMD for the same job. Multi-brand programs can consolidate the work with AED program management software.
7. Documentation and Program Records
Records are how an inspection becomes proof. The AHA guide says to keep all AED records, including prescriptions, warranties, and maintenance schedules, and to store registration information where more than one current staff member can find it. A usable log records:
- Per unit. Model, serial number, location, date installed, and the prescription and medical oversight details on file.
- Per inspection. Date, inspector, status indicator reading, pad and battery expiration dates, and any fault found.
- Per event. Each use, consumable swap, and recall notice, with your response.
The records carry legal weight. Under federal law (42 U.S.C. 238q), an AED owner's immunity does not cover harm caused by a failure to properly maintain and test the device, to notify local emergency response personnel of its placement, or to train an employee or agent who used it. California goes further: Health and Safety Code 1797.196 requires records of maintenance and testing. Some states also require medical oversight and EMS registration. See AED laws and compliance by state, and use AED maintenance logs and what to document for templates.
8. When an AED Reaches End of Life
No manufacturer data sheet prints a retirement date. Most programs plan replacement in an 8 to 10 year window, anchored on warranties: the LIFEPAK CR2 and the HeartSine samaritan PAD 350P and 360P both carry 8-year warranties. Replace sooner for repeated self-test failures after fresh pads and a new battery, discontinued consumables, physical damage, or an unresolved recall. The full list is in When to Replace Your Old AED: 7 Warning Signs.
The FDA classifies AEDs that are not on its approved list as legacy devices and recommends keeping them available until an FDA-approved AED replaces them. An empty cabinet protects no one.
Treat replacement as a purchase decision, not a like-for-like swap. The complete AED buyer's guide, new versus refurbished AEDs, and AED warranty comparison cover the choices, and the AED machines from every major brand shows current models, and our AED value packages cover a complete program.
When the old unit comes out of service, record its serial number and tell the manufacturer so notices stop going to it. ZOLL classifies a scrapped AED as electronic waste and sends electrodes to an electronics recycler because most contain circuit boards, and the EPA says lithium batteries can be managed as universal waste, with state rules varying. How to dispose of an old defibrillator walks through each step.
9. Recall History and How to Check
The FDA regulates AEDs as Class III devices, and manufacturers report any correction or removal involving a health risk to the FDA within 10 working days. A recall does not always mean pulling the unit. Remedies range from a check or software fix to a free replacement, and the notice states which. Three FDA and manufacturer records show the range:
| Year | Device | Issue | Owner action |
|---|---|---|---|
| 2018 | Philips HeartStart FRx, OnSite, and Home built September 2002 to February 2013 (660,519 units). FDA Class 2, Z-1913-2018. | Faulty resistor, with the risk of no shock being delivered. | Philips notified owners by letter. Owners check status at 800-263-3342. |
| 2022 | Philips adult (M5071A) and infant/child (M5072A) SMART Pads cartridges for HeartStart OnSite and Home. | Gel can separate from the backing when the pads are peeled. | Keep the AED in service, store a spare cartridge, follow voice prompts. Philips planned updated pads. |
| 2025 | ZOLL Powerheart G5, models G5S-01C, G5S-02C, G5S-10C, G5S-11C (73 units). FDA Class 2, Z-1320-2025. | Self-test can fail after prolonged exposure to extreme temperature or humidity. | Monitor self-test results, store at 0 to 50 degrees Celsius and 10 to 95 percent humidity, acknowledge the notice to ZOLL. |
Checking your own AED takes five steps:
- Register the AED with its manufacturer so notices come to you.
- Search the FDA medical device recall database by product name or manufacturer, and review the FDA's weekly Enforcement Report.
- Match your serial number, printed on the label (on the back of Philips units), against any affected range.
- Call the manufacturer if a notice never arrived or the status is unclear.
- Log the notice, your response, and the date. Report device problems to the FDA MedWatch line at 1-800-332-1088.
The AED recall history and how to check your model article lists every major recall by brand.
10. Frequently Asked Questions
How often should an AED be inspected?
At least monthly. The AHA guide assigns a liaison to perform monthly AED checks, and ZOLL and Defibtech both recommend monthly inspections. California sets a floor of every 90 days, which a monthly routine already meets. High-traffic and outdoor units benefit from a weekly walk-by (see Section 2).
What is the average lifespan of an AED?
Most programs plan on 8 to 10 years. No manufacturer prints a retirement date, but warranties run up to 8 years, as on the LIFEPAK CR2 and HeartSine samaritan PAD 350P and 360P. Pads and batteries are replaced several times within that window.
Who is responsible for AED maintenance in the workplace?
The organization that acquired the AED. Best practice, per the AHA guide, is to name a program coordinator and a liaison who performs monthly checks, orders supplies, keeps records, and checks recalls, with a backup for absences. Federal law ties the owner's liability protection to proper maintenance and testing.
What are the legal requirements for AED maintenance in the US?
Federal law sets no maintenance schedule, but 42 U.S.C. 238q conditions owner immunity on maintaining and testing the device and notifying local emergency response personnel. States add specifics. California requires maintenance per manufacturer guidelines, testing at least biannually and after each use, inspection every 90 days, EMS notification, and records. OSHA does not require AEDs, and its best practices guide is advisory. See Section 7 for your state's rules.
How do you check if an AED is functional?
Read the status indicator: ready means the last self-test passed, while a red light or chirping means a fault. Then confirm the pads are sealed and in date, the battery is in date, and the case and cables are undamaged. Do not power the AED on to test it.
Are there legal penalties for failing to maintain an AED?
Penalties vary by state, and the larger exposure is civil liability. Federal immunity for AED owners does not apply to harm caused by a failure to properly maintain and test the device, and gross negligence also removes it. California's statute states that a violation is not subject to penalties under Health and Safety Code 1798.206. This is general information, not legal advice.
How do I manage different maintenance schedules if I have multiple AED brands?
Keep one register listing every unit's model, serial number, battery and pad expiration dates, and last inspection. Pads run 2 to 5 years and batteries 4 to 7 years by model, so key your calendar to each unit's earliest expiration date. Consolidating brands cuts the cycles you track.
What is the total cost of ownership for an AED over 10 years?
A single-unit deployment runs approximately $4,000 to $5,500 over 10 years, and ongoing upkeep (pads, battery, refreshers) averages $85 to $150 per year excluding the device. Four scenarios, including a $4,240 mid-range example, are in AED total cost of ownership.
Can I perform the inspections myself, or does a certified technician need to do it?
You can. Visual inspections, cleaning, and replacing pads and batteries are owner tasks per manufacturer guidance, and the AHA guide assigns monthly checks to staff liaisons, not technicians. Send the unit to the manufacturer if the self-test still fails after fresh pads and a new battery.
What happens to my liability if the AED pads are expired but the device is used?
In an emergency, call 911 and use the AED. Expired pads reduce effectiveness because the gel dries out, but withholding the shock guarantees no benefit. The legal exposure sits with the lapse itself: under 42 U.S.C. 238q, an owner's immunity does not cover harm caused by failure to properly maintain the device. Replace pads before they expire and consult counsel on your state's rules.
Is AED maintenance software worth the investment for a small business?
For one AED, a recurring calendar reminder tied to the printed dates is enough. Multi-location programs gain the most, at roughly $50 to $200 per AED per year. See AED program management for options.
How do environmental factors like cold weather affect AED batteries and pads?
Operate and store within the manufacturer's range, 0 to 50 degrees Celsius (32 to 122 degrees Fahrenheit) for the ZOLL AED 3, LIFEPAK CR2, and HeartSine 350P and 360P. HeartSine requires a unit transported below 0 degrees Celsius to return to range for at least 24 hours before use. Outdoor units need a temperature-controlled AED wall cabinet.
What is the process for disposing of old AED batteries and pads safely?
Do not put them in the trash. The EPA says lithium-ion and single-use lithium batteries can be managed as universal waste, and ZOLL recommends sending electrodes to an electronics recycler because most contain circuit boards. State rules vary, so check yours before shipping.
What should I do after the AED is used?
Replace the pads immediately, since pads are single-use. Inspect and clean the AED, check the battery against the manual because some must be replaced after any use, and record the event in your log. Follow your program's reporting steps with medical oversight and EMS.