The eight checks that confirm an AED is rescue-ready, what "ready" looks like on each major brand, and what to do the moment it is not.
The 30-Second Answer
Once a month, confirm eight things: the AED's location, its status indicator, the case and cables, the pads, the battery, the rescue kit, the cabinet and signs, and the temperature of its surroundings. Then sign and date the inspection tag. If the indicator is not showing ready, stop and follow your manufacturer's fault steps in Section 3.
The monthly inspection is the core routine of any complete AED maintenance guide. The American Heart Association's AED implementation guide assigns it to a named liaison, and ZOLL and Defibtech both recommend it. The AED's own self-test covers the electronics, battery, and pads, and AED self-test: what it tests and what it misses is exactly what this checklist covers: placement, access, accessories, and whether anyone is watching the indicator.
What This Article Covers
1. The Monthly AED Inspection Checklist
Use your manufacturer's checklist if it provides one. This list covers the checks the AHA guide, ZOLL, and Defibtech share.
- Location. In its designated spot, visible and unobstructed. The AHA warns against locking an AED in an office or cabinet, where access delays use.
- Status indicator. Read it without opening the case. The ready signal differs by brand (Section 2).
- Case, cables, and connectors. Clean, dry, and free of cracks or loose parts, with cables free of cuts or exposed wire.
- Pads. Present, sealed, and unexpired, plus a spare adult set and pediatric pads where children are present (AHA). Philips pads print the expiration date on the casing.
- Battery. Inside its install-by or expiration date. A LIFEPAK CR2 battery must be installed within 1 year of its manufacture date.
- Rescue kit. Towel, razor, CPR face shield, scissors, and gloves are the usual contents.
- Cabinet and signs. On an alarmed cabinet, confirm the alarm sounds, the door opens easily, and nothing blocks it. The AHA says to inspect signs at least once a year.
- Environment. Philips, HeartSine, ZOLL, and Stryker all rate their current AEDs to operate at 32 to 122 degrees Fahrenheit (0 to 50 degrees Celsius).
Finish by signing and dating the inspection tag. Print this page with your browser's print command, or use the printable monthly AED inspection log (PDF).
Anything inside 12 months of expiring goes on the order list. Replacement batteries are available for the Physio-Control LIFEPAK CR2 4-year lithium battery, the Philips M5070A AED replacement battery (4-year), and the ZOLL AED Plus replacement lithium batteries (10-pack), and Defibtech Lifeline adult AED pads ship ready to swap. Timing details are in AED pad expiration: why it matters and when to replace and the AED battery replacement schedule by brand.
2. What "Ready" Looks Like, Brand by Brand
The indicator is the fastest check, and every brand signals differently. This table summarizes each manufacturer's published behavior.
| AED | Ready | Not ready |
|---|---|---|
| Philips HeartStart FRx and OnSite | Green Ready light blinks | Light off with single chirps and a flashing blue i-button, triple chirps, or no light and no chirp |
| ZOLL AED Plus | Green check mark | Red X, or a beep about once a minute |
| ZOLL AED 3 | Green check mark | Blank window, or a beep once a minute while off |
| Stryker LIFEPAK CR2 | Green Readiness indicator flashes every 6 seconds | Indicator does not flash |
| HeartSine samaritan PAD 350P and 360P | Green light flashes every 5 to 10 seconds | Flashing red, no light, or beeping |
| Defibtech Lifeline | Green indicator flashes (solid when on) | Red, no flash, or dark indicator |
| Cardiac Science Powerheart G3 Plus | Green rescue-ready light | Red light with beeping every 30 seconds |
Indicator designs vary by model year, so confirm against your owner's manual. how to read AED status indicators: lights and beeps covers every signal in detail.
3. If the Indicator Is Not Ready
A non-ready indicator means the last self-test found a problem. Work through these steps in order:
- Read the prompt. On a Philips FRx, press the blinking blue i-button: single chirps point to the pads, battery, or a self-test error, and triple chirps mean call Philips. On a Defibtech, turn the unit on and listen to the voice prompts.
- Check the consumables. On a HeartSine, check the Pad-Pak expiration date first. On a ZOLL AED 3 with no green check, confirm the pad cable is plugged in, replace the pads, then hold On/Off for 5 seconds to run a self-test.
- Replace a low or expired battery. A ZOLL AED 3 beeping once a minute while off needs a new battery pack.
- Still not ready? Remove it from service. Call the manufacturer (Philips at 800-263-3342, ZOLL Technical Service) and move another AED into place or post the nearest alternative.
- Log the fault and the fix.
Call 911 and start CPR first. Philips' manual directs users to press the i-button and follow the prompts even after a serious-fault chirp. Do not abandon a response because an AED shows a fault.
4. After the AED Is Moved, Dropped, or the Cabinet Is Opened
- Moved. Confirm the indicator still shows ready before storing it, notify staff of the new location, and update your EMS registration and records.
- Dropped or damaged. Defibtech's manual says to run a manual self-test after a drop or mishandling. Philips says to run its battery insertion test (remove the battery for 5 seconds, reinstall, wait about a minute) when damage is possible, and not more often than needed because it drains the battery.
- Cabinet opened. Treat an alarm as an inspection trigger: pads still sealed and untampered, kit complete, indicator ready, then re-arm the alarm. If the opening was a rescue, replace the pads before the AED goes back (see AED wall cabinets).
- Moved from cold. HeartSine requires a unit transported below 0 degrees Celsius to return to its 0 to 50 degree range for at least 24 hours before use.
5. Recording the Inspection
Log the date, inspector, indicator reading, pad and battery expiration dates, kit and cabinet condition, and any fault and fix. Philips instructs owners to record each check in the inspection log and to keep the maintenance tag with the unit. The AHA says to keep registration, warranty, and maintenance records where more than one staff member can find them.
The log has legal weight. Under 42 U.S.C. 238q, an owner's immunity does not cover harm caused by failure to properly maintain and test the device, and California's Health and Safety Code 1797.196 requires maintenance and testing records. Templates are in AED maintenance logs and what to document.
6. Frequently Asked Questions
How often should an AED be inspected?
Monthly at minimum, per the AHA guide, ZOLL, and Defibtech. California requires inspection at least every 90 days. HeartSine's manual also calls for a weekly indicator check, which suits high-traffic and outdoor units.
What should be on an AED checklist?
Location, status indicator, case and cables, pads, battery, rescue kit, cabinet and signs, and surrounding temperature, ending with a signed and dated tag. See Section 1.
Do AEDs need annual servicing?
None of the manufacturer manuals reviewed requires annual technician service. Owners run the visual checks, and the AED self-tests automatically. Defibtech states that the program's medical director sets the final maintenance schedule, and some states set testing intervals, so confirm both.
Who is responsible for checking the AED in a workplace?
The organization that owns it. The AHA guide assigns monthly checks to a named liaison. Name a backup so a vacation never skips a month.
What happens if an AED fails its self-test, or the indicator turns red?
The indicator changes (red, blank, or no longer flashing) and many models chirp or beep. Read the prompt, check pads and battery, and remove the unit from service if it stays not ready. Full steps are in Section 3.
Is there a legal requirement to keep a written log of AED checks?
No federal statute sets a log format, but California requires records of maintenance and testing, and federal immunity under 42 U.S.C. 238q depends on proper maintenance and testing. A log is how you prove both. State rules are summarized in AED laws and compliance by state.
Can I use the same checklist for different brands like ZOLL, Philips, or LIFEPAK?
The eight checks are the same. The ready signal, fault sounds, and expiration locations differ, so add a one-line note per unit using the table in Section 2.
How do I identify if the battery is low if the indicator light isn't clear?
Listen. A ZOLL AED 3 beeps once a minute while off and says CHANGE BATTERY at power-up. A Philips chirps while the i-button flashes. A HeartSine gives a low-battery warning with about ten shocks left. A Defibtech says "Battery low" at power-up.
What is the easiest way to store these records for a future audit?
Keep the signed tag on the unit and a master log with its registration, warranty, and prescription records, in a place several staff can reach. Photograph each tag monthly if your records live in a shared folder.
Are digital apps better than paper checklists?
For one AED, a paper tag works. For several, software adds expiration reminders: Stryker's wireless LIFEPAK CR2 emails the administrator when a battery or electrode date approaches, and AED program management software covers mixed fleets.
Who is liable if the AED fails during an emergency because of a missed check?
Federal law protects users who act, but an owner's immunity under 42 U.S.C. 238q does not cover harm caused by failure to properly maintain and test the device. Liability depends on the facts and your state. This is general information, not legal advice.
Do I need to check the AED after it has been moved or the cabinet has been opened?
Yes. Confirm the indicator, pads, and kit, and run the manufacturer's self-test if the unit was dropped. See Section 4.