AED Temperature Range: Heat and Cold Limits by Model

Most AEDs run between 32 and 122 degrees Fahrenheit. A few have a narrower standby range, and a hot car or a freezing garage can push any AED out of bounds. Here are the limits by model, what the beeps mean, and how to protect your unit.

Reading time: 9 minutes Last updated: October 2026 Reviewed by: Kyle Nowack, BLS Instructor Trainer

The 30-Second Answer

The AED temperature range for most current models is 0 to 50 degrees Celsius (32 to 122 degrees Fahrenheit). Philips HeartStart OnSite and HS1 units are the exception: their standby range is 10 to 43 degrees Celsius (50 to 109 degrees Fahrenheit).

Store the AED indoors, or in a cabinet built for the climate. Never leave it in a parked car. See the AED temperature range by model, then read what to do after heat or cold exposure.

1. AED Temperature Range by Model

These figures come from manufacturer specification sheets and manuals. Your own manual wins if it differs.

AED Operating range Standby or storage Worth knowing
Philips HeartStart FRx AED 0 to 50 C (32 to 122 F) Standby 0 to 50 C (32 to 122 F) Short use down to minus 20 C (minus 4 F) is allowed for 20 minutes or less after a rapid change from 20 C.
Philips HeartStart OnSite AED and HS1 0 to 50 C (32 to 122 F) Standby 10 to 43 C (50 to 109 F) The narrowest standby range here. Standby humidity is 10 to 75 percent.
ZOLL AED Plus defibrillator 0 to 50 C (32 to 122 F) Check the manual Humidity 10 to 95 percent, non-condensing.
ZOLL AED 3 defibrillator 0 to 50 C (32 to 122 F) Device storage minus 30 to 70 C (minus 22 to 158 F) Battery 0 to 50 C. Store a spare battery up to 2 years at about 23 C before installing.
Defibtech Lifeline AED defibrillator 0 to 50 C (32 to 122 F) Standby 0 to 50 C (32 to 122 F) Tolerates minus 20 C (minus 4 F) for one hour from room temperature. Humidity 5 to 95 percent.
Physio-Control LIFEPAK CR2 AED 0 to 50 C (32 to 122 F) With battery and electrodes: minus 30 to 60 C (minus 22 to 140 F) for one week at most. Long-term: 15 to 35 C (59 to 95 F). Performance figures assume two hours at operating temperature first.
HeartSine samaritan PAD 350P and HeartSine samaritan PAD 360P 0 to 50 C (32 to 122 F) Standby and transport 0 to 50 C (32 to 122 F) After transport below 0 C, allow at least 24 hours at 0 to 50 C before use.
Cardiac Science Powerheart G5 AED 0 to 50 C (32 to 122 F) Keep within 0 to 50 C and 10 to 95 percent humidity Humidity outside the limit is linked to error EC 501. See the AED recall history and how to check your model.

The pattern is simple. The operating range is the same almost everywhere. The standby range is where models differ, and the Philips OnSite and HS1 is the one that catches owners out.

Many web guides say 32 to 122 degrees for all units. That puts an OnSite in a cold garage or a hot attic that its standby range does not allow.

2. Operating, Standby and Storage Limits Explained

Manufacturers print three kinds of limits. They are not the same thing.

  • Operating range. The air temperature in which the AED is built to deliver a rescue.
  • Standby range. The range for an AED that is switched off and running its daily self-tests while it waits. This is the one that matters for a wall cabinet.
  • Storage or transport range. A wider range for a unit that is boxed or moved, often with a time limit. The LIFEPAK CR2 limits its wide range to one week.

Humidity has its own limit, and it is non-condensing. Moving an AED from a freezing space into a warm humid room can fog it. Let it warm up in a dry place before you judge its readiness.

For how water and dust ratings differ, see AED IP ratings compared: which AEDs survive dust and water.

3. What Heat Does to AED Pads and Batteries

Heat works slowly on the parts that expire, so the damage is easy to miss.

  • Pad gel dries out. The gel is water-based. Heat speeds evaporation, and pads stored above their limits expire early. See AED pad expiration: why pads expire and when to replace.
  • Batteries lose charge faster. All batteries self-discharge. Heat in standby speeds it, which shortens battery life. See the AED battery replacement schedule by brand.
  • Out-of-range storage can cut shelf life. HeartSine states that operating or storing the unit outside its ranges can cause malfunction or reduce Pad-Pak shelf life.

Can an AED be left in a hot car?

No. A closed car heats fast. In a study on 16 days with outside temperatures from 72 to 96 degrees Fahrenheit, the cabin rose an average of 19 degrees in 10 minutes and 43 degrees in 60 minutes. Cracking the windows changed the result by less than 3 degrees.

Do the math for an 80-degree day. After 60 minutes the average cabin reaches about 123 degrees, which is above the 122-degree limit for nearly every AED here. On a 95-degree day it passes 122 in under 30 minutes.

Interiors vary, since the study used one dark sedan. Treat the numbers as a warning, not a forecast.

If your program requires a vehicle AED, carry it in and out with you. Do not leave it in the trunk overnight. Look at the AED cases collection and the AED rescue kits collection for carrying options.

4. What Cold Does to AED Pads and Batteries

Cold hits the battery first.

  • Battery power drops. Distributor guidance states batteries drain faster in cold. The ZOLL AED 3 and Powerheart G5 batteries are rated for use at 0 to 50 degrees Celsius.
  • Pad gel can freeze. The gel is water-based, and distributors warn it can freeze below 32 degrees Fahrenheit. No manufacturer we reviewed says whether a freeze permanently damages the pads.
  • The self-test can fail. Distributor guidance says AEDs left in the cold commonly start beeping, so a cold unit can flag a fault.

Because the damage is not documented, use a safe rule. Treat pads that were below the range as suspect. Replace them if the gel looks dry or cracked or the pouch is damaged. HeartSine says not to use pads that are dry, unsealed or damaged.

Short cold exposure is allowed on some models. The Philips FRx permits operation down to minus 20 degrees Celsius for 20 minutes or less, and the Defibtech Lifeline allows one hour at minus 20. That is for a rescue, not for storage.

Do AED pads go bad if they freeze?

They can. There is no published figure for how many freeze cycles pads survive. Replace any pads that froze if you see dry gel, cracks or a torn pouch, or if the manufacturer tells you to. Order replacements from the AED pads collection.

5. AED Temperature Alarm Meaning: Beeps, Lights and Messages

AEDs rarely say "temperature" in plain words. A chirp or a red indicator is a general fault, and temperature is one cause among several.

Signal What it means
Philips FRx: blinking Ready light The last self-test passed and the AED is ready.
Philips FRx: Ready light off, chirping, i-button flashing Self-test error, a pad problem, the Infant/Child Key left in, or a low battery. Press the i-button for instructions. The manual page does not tie this to temperature.
Powerheart G5: red indicator, beeping A fault was found. Distributor guidance says an out-of-range G5 can beep and its diagnostic screen can show "Temp Too Cold".
Powerheart G5: red X with a service message ZOLL says to return the device for service when "SERVICE REQUIRED" appears. Error EC 501 is most likely linked to humidity outside the limit.

For each brand's lights and tones, read AED status indicators: decoding lights and beeps and AED self-test: what it checks and what failures mean.

What should I do if my AED shows a temperature warning?

  1. Check the temperature of the spot. Compare it with your model's standby range in the table above.
  2. Move the AED to a room at normal temperature. Keep it dry.
  3. Give it time. HeartSine requires at least 24 hours at 0 to 50 degrees Celsius after transport below 0. One distributor suggests at least 30 minutes for a beeping unit. Use the longer figure when in doubt.
  4. Read the indicator again. If it returns to ready, log the event. If it stays red, replace the battery and pads if they are due, then call the manufacturer.
In an Emergency

If someone collapses, call 911 and use the nearest AED, even one that was in a hot or cold spot. A working AED that has been exposed is better than no shock. Follow the voice prompts and start CPR.

6. What to Do After Heat or Cold Exposure

Use the same sequence each time. It takes five minutes.

  1. Bring the AED to normal temperature. Do it gradually and keep it dry.
  2. Check the status indicator. Do not power the AED on to test it. The self-test already reports its state.
  3. Inspect the pads. Look for a swollen, torn or leaking pouch. If you can see the gel, look for dry or cracked areas.
  4. Check the dates. Pad and battery dates assume storage inside the range. Replace early if the unit sat outside it for days.
  5. Record it. Write the date, the temperature and what you found in your AED maintenance logs: what to document and why.
  6. Call the manufacturer if the fault stays. Have the model and serial number ready.

For the full monthly routine, use the AED inspection checklist: monthly, quarterly and annual. The complete picture is in the complete AED maintenance guide.

7. How to Protect an AED from Extreme Temperatures

Match the storage to the location.

Location Risk What to do
Climate-controlled lobby or office Low Keep it out of direct sun and away from vents and heaters.
Garage, warehouse, loading dock High in summer and winter Use a temperature-controlled cabinet, or relocate the AED to a controlled room.
Outdoor public site Highest Use a cabinet made for outdoor use, in shade. Choose a model with a high IP rating.
Vehicle Very high Carry it with you. Do not leave it in a parked car or the trunk.
Shipping or transport Brief but severe Follow the transport range. For HeartSine, allow 24 hours to return to range.

How do you protect an AED from extreme temperatures?

How often should I check the temperature of my AED storage area?

No manufacturer sets an interval. The practical step is to check at every monthly inspection and after any heat wave, cold snap or power failure. A min/max thermometer in the cabinet shows the extremes between checks.

Mark the check on the Avid Safety CPR AED inspection tag so the next person sees it.

Buying for a tough climate

If your site is hot, cold or exposed, choose the AED and the storage together. The IP rating helps with dust and water, not temperature, so check both.

Compare units in the AED machines collection, or read the AED buyer's guide. For ongoing checks across several sites, see AED program management.

8. Frequently Asked Questions

What is the recommended storage temperature for an AED?

Most AEDs are rated for standby at 0 to 50 degrees Celsius (32 to 122 degrees Fahrenheit). The Philips OnSite and HS1 standby range is 10 to 43 degrees Celsius (50 to 109 degrees Fahrenheit). See the AED temperature range by model.

Can an AED be left in a hot car?

No. A study found car interiors rise an average of 19 degrees Fahrenheit in 10 minutes and 43 degrees in an hour. On an 80-degree day that passes the 122-degree limit within an hour.

Do AED pads go bad if they freeze?

Possibly. The gel is water-based and can freeze below 32 degrees Fahrenheit, and no manufacturer publishes how a freeze affects it. Replace pads with dry or cracked gel or a damaged pouch, and follow your manufacturer.

Will freezing temperatures cause the gel on the pads to dry out?

Heat and age dry gel, because they evaporate its water. Freezing is a separate risk, since the gel is water-based. No manufacturer publishes how a freeze affects it, so check the pouch and the gel, and see AED pad expiration: why pads expire and when to replace.

What happens to an AED battery in extreme cold?

It drains faster and delivers less. The batteries we reviewed are rated for use at 0 to 50 degrees Celsius, so a battery below 0 is outside its rating. Warm it indoors and check the indicator. See what cold does to AED pads and batteries.

How do you protect an AED from extreme temperatures?

Store it in a climate-controlled space, or in a cabinet made for hot or cold sites. Keep it out of direct sun, vehicles and unheated buildings. See how to protect an AED from extreme temperatures.

How often should I check the temperature of my AED storage area?

No manufacturer sets an interval. Check it at every monthly inspection and after extreme weather, and log the result.

What should I do if my AED shows a temperature warning light?

Move it to a normal-temperature room, keep it dry, and give it time to return to range. HeartSine requires at least 24 hours after transport below 0 degrees Celsius. If the red indicator stays on, call the manufacturer. See AED temperature alarm meaning.

How do I know if the heat has ruined my AED pads?

Check for a swollen or torn pouch and for dry or cracked gel. Replace the pads if the AED sat above its range or if the pouch is damaged. You cannot confirm gel health from the outside of a sealed pouch, so replace early when in doubt.

Can I store an AED outside?

Only in a cabinet made for outdoor use, in shade, with a unit rated for the conditions. The Philips OnSite and Home models have an IP21 rating and are not for outdoor storage. See AED IP ratings compared.

Does humidity matter for AED storage?

Yes. Manufacturers list non-condensing humidity limits, such as 10 to 95 percent for the ZOLL AED Plus and Powerheart G5. ZOLL linked G5 error EC 501 to humidity outside the limit.

See our full shipping policy and return policy before ordering, or call (855) 286-6384 to match an AED, pads and battery to your climate.

Kyle Nowack, BLS Instructor Trainer and AED Program Specialist at Total AEDs
Written and Reviewed By

Kyle Nowack

BLS Instructor Trainer, AED Program Specialist

Kyle Nowack is a certified American Heart Association BLS Instructor Trainer and AED program specialist at Total AEDs. His work focuses on helping workplaces, schools, and public-access programs build defensible AED deployments, from initial device selection through compliance documentation and responder training. Kyle has trained hundreds of workplace and school responders on CPR and AED use across the U.S.

Add Comment

?
TOP
0 Items
Logo